Water quality parameters are monitoring using in-situ probes.  The probes connected to multiprobe sondes.  The probes measure selected parameters, converting these measures in electrical singles that are processed by the sonde.  Data are then sent up a cable that is connected to the computer contained with the buoy’s hull.

YSI_Sonde

H2O_Sonde

xmiss

YSI 6600 Sonde

Hydrolab DataSonde 4a

Wetlab Transmissometer

Both YSI 6600 and Hydrolab Datasonde 4a sondes are being utilized by this project.

Refer to the tables below (the upper table refers to the YSI 6600 probes, while the lower table refers to the Hydrolab Datasonde 4a) to obtain the model number, sensor type, range, accuracy and resolution of the various probes.

 

YSI 6600: Otisco Lake and Lake George and Skaneateles Lake (2002 only)

Parameter

Model number

Sensor Type

Range

Accuracy

Resoluti on

depth

integrated

Strain gauge

0 to 61 m

± 0.12 m

0.001 m

water temperature

6560

thermistor

-5 to 45 ºC

±0.15 ºC

0.01 ºC

specific conductance

4 electrode cell with autoranging

0 to 100 mS/cm

±0.5% of reading or 1 uS/cm

1 uS/cm

dissolved oxygen

6562

Rapid Pulse, Clark type, polarographic

0 to 50 mg/L

±2% of reading or 0.2 mg/L

0.01 mg/L

pH

6565

Glass combination electrode

0 to 14 units

±0.2 units

0.01 units

redox potential

Platinum button

-999 to 999 mV

±20 mV

0.1 mV

turbidity

6026

Optical, 90º scatter with wiper

0 to 1000 NTU

±5% of reading or 2 NTU

0.1 NTU

chlorophyll

6025

Optical fluorescence with wiper

0 to 400 ug/L

manufacture doesn’t report

0.1 ug/L

Hydrolab Datasonde 4a Skaneateles Lake (2003)

Parameter

Model number

Sensor Type

Range

Accuracy

Resoluti on

depth

integrated

Strain gauge

0 to 100 m

± 0.3m

0.1 m

water temperature

 

thermistor

-5 to 50 ºC

±0.1 ºC

0.01 ºC

specific conductance

 

4 electrode cell with autoranging

0 to 100 mS/cm

±0.5% of reading or 1 uS/cm

1 uS/cm

turbidity

 

Optical, 90º scatter with shutter

0 to 100 NTU

±2.6% of reading or 2 NTU

0.1 NTU

transmissometry

Wetlabs

660 nm 25 cm pathlength

0 to 100%

0.2%

0.1%

 